Matters needing attention
ZDA series of scattered light turbidity meter is a precision measuring instrument which is combined with photoelectric. It is necessary to read the instruction manual carefully before operation and get accurate measurement results.
1, the use of the environment must meet the working conditions.
2, the measurement tank must be clean and dry for a long time, no dust, no need to cover the cover.
3, humid climate, must be extended to boot time.
4, the solution should be carefully poured into the sample bottle, to prevent the formation of bubbles, affecting the accuracy of measurement.
5, replace the sample bottle or after repair must be re calibration.
6, non professional maintenance engineer, do not open the instrument for maintenance.
Working conditions
1. The instrument should be placed in the drying room, use temperature of 5 DEG C – 35.
2. When used, should be placed on the flat surface, and avoid vibration.
3. Indoor lighting should not be too strong, and avoid direct sunlight irradiation.
4. Try to keep away from the high intensity magnetic field, electric field and high frequency electric equipment.
5. Avoid high temperature close to the instrument.
6. The power supply of the instrument: 220V± 22V, 50± 1Hz, and must have a good grounding wire.
Measurement preparation
1, open the instrument power switch;
2, do not fall wool soft cloth to wipe the water stain on the sample bottle and fingerprints, such as not easy to clean the cleaning agent can be soaked, and then rinse with water clean;
3, ready for school pocket zero turbidity water and preparing for the calibration of formazine turbidity standard solution
4, with a clean container to collect a representative sample;
Operation steps(see Appendix flow chart)
Measurement solutionturbidityThe instrument must be calibrated before.
1 Correction
Boot into the main interface menu window,“ switching ” key select “ correction ” option.At this time the instrument display into0000NTU turbidity liquid, open the machine above the sliding cover panel,Will be installedYesZero turbidity waterSample bottleMeasurement pool,Cover with sliding coverTo be stabilized by “ confirm ” key. In order to put in the correct solution. Press “ confirm ” key end. Enter the main menu option. Or press “ switch ” key forced exit, exit to save the corrected data. Enter the main menu interface.
Note: every time to wipe cleanSample bottle wallTraces of water and fingerprints,Should not leave any traces, so as not to affect the measurement accuracy. every timeSample bottleThe direction and the position of the measuring pool are as consistent as possible.Please use the same sample bottle when comparing and measuring, so as to reduce the inaccuracy of the data value caused by the error of the sample bottle.
2,turbidity measure
Main interface menu window,By “ switch ” key select “ measure ” option.
Will be measuredsamplePour cleanSample bottleIn, will be installedYesTo be measuredsampleSample bottleMeasurement pool,Cover with sliding cover,The numerical display is the instrument to be checked.Turbidity value.
3,ProtectDeposit,extractUsage method
In the measurement process, the measurement can be carried outPreservation. Press“ save ”Key can be stored up to10 sets of data. Sign outPreservationPress“confirm”Key, back to the status of the measurement.
PreservationData can be queried. Press“ extraction ”Key can. Sign outextractPress“confirm”Key, back to the status of the measurement.